Skip to main content

Table 3 Agreement of in-person auscultation and Stemoscope tele-auscultation results among three cardiologists

From: Real-world evaluation of the Stemoscope electronic tele-auscultation system

 

Cardiologist 1

Cardiologist 2

Cardiologist 3

In-person

In-person

In-person

Abnormal

Normal

Abnormal

Normal

Abnormal

Normal

Tele-auscultation

 

Abnormal

153

14

150

16

141

18

Normal

6

85

12

80

11

88

Agreement

92%

89%

89%

P value

0.27

0.57

0.26

Kappa value

0.77

0.77

0.83