Fig. 1

From: In vivo experimental validation of detection of gastric slow waves using a flexible multichannel electrogastrography sensor linear array

Fig. 1

Benchtop test results of two opamps for EGG recordings in response to synthetic sinusoidal signals at 1, 3, 5, 7 and 9 cpm. A TLC272BID and (B) OPA2325 had a similar high power of 70 dB at 3 cpm, low mains interference of − 4 dB, but the harmonic interference at 100 Hz in TLC272BID (− 14 dB) is higher than it in OPA2325 (− 20 dB)