Skip to main content

Table 1 Biomechanical test results of three fixation methods

From: Biomechanical comparison of the femoral neck system versus InterTan nail and three cannulated screws for unstable Pauwels type III femoral neck fracture

Group

Axial stiffness (N/mm)

A-P bending stiffness (N/mm)

Torsional stiffness (Nm/°)

Overall deformation (mm)

Local deformation of implant holes (mm)

Group A

423.71 ± 18.69

227.46 ± 26.07

0.32 ± 0.01

2.45 ± 0.25

1.2 ± 0.1

Group B

485.63 ± 13.27a

410.08 ± 50.37a

0.31 ± 0.03

2.02 ± 0.10a

0.81 ± 0.08a

Group C

375.75 ± 14.32a,b

145.2 ± 14.83a,b

0.22 ± 0.02a,b

3.39 ± 0.23a,b

2.18 ± 0.07a,b

F

74.781

96.280

45.222

70.550

435.023

P

 < 0.001

 < 0.001

 < 0.001

 < 0.001

 < 0.001

  1. aP < 0.01, vs. Group A; bP < 0.01, vs. Group B
  2. Values are presented as mean ± standard deviation
  3. F and P values are obtained by one-way analysis of variance
  4. Different letters (a, b) indicate significant differences in the same column at a P < 0.01 by Bonferroni post hoc test