Table 5 Reduction of the Young’s Modulus of the conditioned samples compared to the dry samples for the dog bone, tube, and necked tube specimens at a test speed of 50 mm/min at each temperature level

From: Temperature-dependent tensile properties of polyamide 12 for the use in percutaneous transluminal coronary angioplasty balloon catheters

Sample 23 °C \(37\) °C \(50 ^\circ \mathrm{C}\) 80 °C 100 °C
Dog bone − 51.7 [%] − 38.6 [%] − 23.4 [%] − 5.9 [%] − 2.9 [%]
Tubes − 7.0 [%] − 18.6 [%]
Necked Tubes − 8.6 [%] − 19.1 [%]