Table 4 Median values and IQR of the Young’s Modulus \(E\) and proportional limit \({\sigma }_{\mathrm{p}}\) of the tensile tests on tube specimens

From: Temperature-dependent tensile properties of polyamide 12 for the use in percutaneous transluminal coronary angioplasty balloon catheters

Temperature [°C] State \({E}_{1\mathrm{mm}/\mathrm{min}}\) [MPa] (IQR) \({E}_{50\mathrm{mm}/\mathrm{min}}\) [MPa] (IQR) \({\sigma }_{p}\) [MPa] (IQR)
23 Dry 1209 (54.2) 1303.8 (69.7) 23.2 (0.9)
  Cond 1159.2 (85.7) 1191.6 (68.58) 19.53 (1.0)
37 Dry 1123.1 (54.7) 1172.0 (109.6) 17.68 (1.0)
  Cond 776.8 (114.8) 948.0 (75.6) 15.4 (1.1)