Skip to main content

Table 2 Results of the denoising performance analysis on the sequence with the moving rectangle

From: Toward a priori noise characterization for real-time edge-aware denoising in fluoroscopic devices

Sequences a b F \({a}_{est}\) \({b}_{est}\) \(MSE\left({S}_{filt},{S}_{fil{t}_{err}}\right)\) \(\Delta FWHM\)
1–7 2 0 25 1.979 3.007 0.090284 0.001851
2 0 10 1.937 7.186 0.199485 0.002836
8–14 2 0 25 1.425 74.866 2.276551 0.041098
2 0 10 0.984 131.379 5.324343 0.256335
  1. The dissimilarity scores between the sub-optimal and optimal filtered sequences are reported, along with the corresponding noise parameters estimates, the number of frames (F) used in the estimation, and the actual noise parameters values