Table 1 System hardware specifications

From: A real-time system for selectively sensing and pacing the His-bundle during sinus rhythm and ventricular fibrillation

96-channel preamplifier—(built inhouse)

 Preamplifier gain

162x

 Input-referred noise

100 µV

 Input impedance-between reference and each electrode

1010Ω

 Frequency bandwidth

2.14 kHz

 Bandpass center frequency, Q-factor

28.88 Hz, 0.013

 SNR

69

 Lower cut-off frequency, Q-factor

0.39 Hz, 1

 Upper cut-off frequency, Q-factor

2.14 kHz, 0.5

96-channel real-time data-acquisition, recording and current routing hardware—(National Instruments, TX hardware modules assembled inhouse)

 Max sampling rate for each of 96 channels

6.25 kHz

 Analog output channel configured for voltage to current converter

1 channel

 Time to configure current source to a chosen channel

1.89 s

 Time to trigger the switch after current channel configured

20 ms

Stimulator—(custom built by rdmApps, CO, USA)

 Stimulator

Unipolar, referenced to ground

 Voltage to current conversion

±10 V to ± 10 mA

 Output current range

±10 mA

 Output frequency

0–10 kHz

 Max impedance

10 kΩ