Skip to main content

Table 1 A summary of the mean error values \(\delta_{k}\) and the minimum, maximum values and standard deviations of the error \(\delta_{k}\) obtained for all the compared methods: Sobel, Roberts, Canny and the author’s (proposed) method of outer corneal edge detection in images from the Corvis® ST tonometer

From: A new method for detecting the outer corneal contour in images from an ultra-fast Scheimpflug camera

Method

δk (%)

δk(min) (%)

δk(max) (%)

std (%)

Proposed

0.16

0.09

3.62

0.19

Sobel

3.43

0.25

42.12

6.21

Roberts

5.78

0.17

61.67

9.19

Canny

1.26

0.53

50.70

3.11