From: High-frequency irreversible electroporation for cardiac ablation using an asymmetrical waveform
5 animals (model M) | ||
---|---|---|
Symmetrical | Asymmetrical (β = 2.31) | |
Positive voltage peak (V) | 600 | 544 |
Negative voltage peak (V) | − 600 | − 221 |
Frequency (kHz) | 167 | 167 |
Number of wave trains | 10 | 10 |
Interval (ms) between wave trains | 400 | 400 |
Duration (ms) of single wave train | 2 | 2 |
Delivered energy | 10 × 20 J | 10 × 20 J |
Energy during one single wave period T | 100% | 100% |
Charge during one single wave period T | 100% | 92% |