Table 3 Performance measures of tested heart beat detector on SCG signals p001–p020

From: Comparison of HRV indices obtained from ECG and SCG signals from CEBS database

Signal

TP

FP

FN

Beats

Se

PPV

p001

317

11

8

325

0.975

0.966

p002

308

0

0

308

1.000

1.000

p003

95

254

253

348

0.273

0.272

p004

324

2

1

325

0.997

0.994

p005

181

240

184

365

0.496

0.430

p006

139

226

224

272

0.176

0.381

p007

273

1

0

273

1.000

0.996

p008

479

0

1

480

0.998

1.000

p009

313

6

3

316

0.991

0.981

p010

234

75

74

308

0.760

0.757

p011

251

88

86

337

0.745

0.740

p012

317

81

87

404

0.785

0.796

p013

358

0

1

359

0.997

1.000

p014

344

1

0

344

1.000

0.997

p015

328

3

1

329

0.997

0.991

p016

352

0

0

352

1.000

1.000

p017

363

2

2

365

0.995

0.995

p018

400

0

0

400

1.000

1.000

p019

316

0

0

316

1.000

1.000

p020

326

15

12

338

0.964

0.956

Total

6018

1005

937

6864

0.877

0.857