Table 1 Performance measures of tested heart beat detector on SCG signals b001–b020

From: Comparison of HRV indices obtained from ECG and SCG signals from CEBS database

Signal

TP

FP

FN

Beats

Se

PPV

b001

279

22

19

298

0.936

0.927

b002

308

0

0

308

1.000

1.000

b003

121

187

226

347

0.351

0.395

b004

323

2

1

325

0.997

0.994

b005

139

226

224

364

0.385

0.383

b006

309

2

0

309

1.000

0.994

b007

272

1

0

272

1.000

0.996

b008

480

0

1

480

0.998

1.000

b009

310

6

3

313

0.990

0.981

b010

234

75

74

309

0.761

0.758

b011

251

86

86

338

0.746

0.741

b012

317

81

86

403

0.787

0.797

b013

358

0

1

359

0.997

1.000

b014

345

1

0

345

1.000

0.997

b015

329

3

1

330

0.997

0.991

b016

352

0

0

352

1.000

1.000

b017

363

2

2

365

0.995

0.995

b018

400

0

0

400

1.000

1.000

b019

316

0

0

338

1.000

1.000

b020

338

15

12

338

0.965

0.956

Total

6137

711

736

6873

0.893

0.896