Table 6 The existing ACDO devices and their specifications

From: Continuous distraction osteogenesis device with MAAC controller for mandibular reconstruction applications

Refs. Year Mechanism Distraction accuracy (μm) Distraction step error (μm) Operated distraction rate (mm/day) Maximum travel (mm) Distraction force (N) Total size (mm)
[60] 1999 Motor-based 40 0.5 1 13.6
[51] 2000 Hydraulic 1.5 30 to 50
[40] 2004 Motor-based 1000 20 1 15 70 60
[62] 2005 Hydraulic 1 16 20
[25] 2008 Motor-based 80 0.9 10 19 55
[44] 2009 Hydraulic 10 86 1 25 25 to 40 30 to 100
[26] 2009 Syringe pump 21,000 0.9 15
[35] 2010 Motor-based 600 15 35.6
[24] 2011 Motor-based 200 2000 1.4 7 2.84 35
[63] 2011 Motor-based 0.75 30 3 3
[7] 2013 Hydraulic Average < 500 1.5 12 25 to 40 18
Average < 1000 3
4.5
[5] 2014 Motor-based 300 4 2.4 18
[32] 2015 Hydraulic 3 30 25 to 40
Proposed device Motor-based 0.00763 0.00006 1 22 38 25
0.00006 3
0.00002 5