Skip to main content

Table 2 The positioning accuracy of the system

From: Continuous distraction osteogenesis device with MAAC controller for mandibular reconstruction applications

Drive mode

Rotor stride angle (°)

Shaft stride angle (°)

Positioning accuracy (nm)

Carriage movement (nm/step)

Full-step (1/1)

5.625

0.088

244.14

244.14

Half-step (1/2)

2.8125

0.044

122.07

122.07

Micro-step (1/32)

0.176

0.00275

7.63

7.63