From: Analysis of measurement electrode location in bladder urine monitoring using electrical impedance
Subject no. | VCR (%) | Pearson’s correlation coefficient R | ||||
---|---|---|---|---|---|---|
M1–M2 | M3–M4 | M5–M6 | M1–M2 | M3–M4 | M5–M6 | |
1 | 9 | 26 | 5 | − 0.993 | − 0.913 | − 0.938 |
2 | 4.4 | 14.3 | 2.7 | − 0.847 | − 0.941 | − 0.978 |
3 | 16.6 | 23.9 | 9 | − 0.885 | − 0.978 | − 0.934 |
4 | 13.8 | 16.8 | 10.6 | − 0.927 | − 0.942 | − 0.978 |
5 | 28.6 | 57.6 | 12.5 | − 0.912 | − 0.815 | − 0.945 |
6 | 18.4 | 36.1 | 12.5 | − 0.992 | − 0.945 | − 0.841 |
7 | 13.9 | 32.2 | 8.3 | − 0.920 | − 0.835 | − 0.939 |
8 | 8 | 13.2 | 3.9 | − 0.962 | − 0.958 | − 0.977 |
Mean ± std | 14.1 ± 7.5 | 27.5 ± 14.7 | 8.1 ± 3.8 | − 0.930 ± 0.051 | − 0.916 ± 0.059 | − 0.941 ± 0.045 |