From: Novel dynamic corneal response parameters in a practice use: a critical review
Author | Participants | Parameters | Results |
---|---|---|---|
Lopes et al. [81] | 32 healthy patients | IOP; bIOP; and eight selected DCR parameters | Corvis ST showed good repeatability and reproducibility for unadjusted IOP measurements and biomechanically corrected IOP |
Miki et al. [82] | 48 healthy patients | IOP and 35 DCR parameters available in the Corvis ST tonometer | Repeatability for 22 out of 35 analysed corneal deformation parameters was at a high level (the intraclass correlation coefficient ICC ≥ 0.75) |