From: Non-orthogonal one-step calibration method for robotized transcranial magnetic stimulation
Data group | Algorithm | Error type | Total errors | P n |
---|---|---|---|---|
Simulation | QR24 | Translation | < 0.4 mm | 118 |
< 0.3 mm | None | |||
Rotation | < 0.2° | 10 | ||
< 0.17° | 28 | |||
QR36 | Translation | < 0.4 mm | 7 | |
< 0.3 mm | 14 | |||
Rotation | < 0.2° | 6 | ||
< 0.17° | 22 | |||
Experiment | QR24 | Translation | < 1 mm | 247 |
< 0.7 mm | None | |||
Rotation | < 0.4° | 10 | ||
< 0.35° | 66 | |||
QR36 | Translation | < 1 mm | 9 | |
< 0.7 mm | 27 | |||
Rotation | < 0.4° | 6 | ||
< 0.35° | 57 |