Table 1 Measured values of the stimulation parameters

From: Design of a novel stimulation system with time-varying paradigms for investigating new modes of high frequency stimulation in brain

Parameters

Settings

Measured values (mean ± standard deviation)

|Mean-Setting|/Setting (%)

Intensity (V)

3

3.0112 ± 0.0005

0.37

Frequency (Hz)

100

100.0010 ± 0.0388

0.001

Duration (s)

5

4.9982 ± 0.0038

0.036

Pulse width (ms)

0.1

0.1027 ± 0.0037

2.7

  1. Number of pulse samples is 500