Skip to main content

Table 1 Maximum duration of the effective ventriculoatrial interval in the method of Janoušek et al. [14]

From: R-wave synchronised atrial pacing in pediatric patients with postoperative junctional ectopic tachycardia: the atrioventricular interval investigated by computational analysis and clinical evaluation

MTR (bpm) VAImax (ms) MTR (bpm) VAImax (ms) MTR (bpm) VAImax (ms)
230 190 180 260 130 390
220 200 170 280 120 400
210 210 160 300 110 400
200 220 150 320 100 400
190 240 140 350 90 400
  1. MTR maximum tracking rate, VAI max maximum effective ventriculoatrial interval