Skip to main content

Table 2 Maximum von Mises stress (MPa) of three different fixation modalities

From: A customized fixation plate with novel structure designed by topological optimization for mandibular angle fracture based on finite element analysis

 

Loading I

Loading II

Loading III

Bone

 Type A

66

80

103

 Type B

57

65

56

 Type C

48

44

63

Fixation system (plate and screw)

 Type A

377

332

394

 Type B

149

123

192

 Type C

97

81

116