Table 6 Errors of blood pump flow rate for the two types of dialysis machines at different time points under different settings

From: In vivo and in vitro performance of a China-made hemodialysis machine: a multi-center prospective controlled study

Test time

Blood pump flow rate (mL/min)

Actual error (ml)

SWS-4000

Fresenius 4008

Week 2

100

69.6 ± 128.7

41.8 ± 78.2

200

146.9 ± 212.3

79.5 ± 76.1

300

307.5 ± 320.5

86.0 ± 86.5*

400

385.4 ± 451.8

287.8 ± 334.3

Week 13

100

80.4 ± 118.1

20.6 ± 37.6

200

175.0 ± 191.3

118.8 ± 155.1

300

326.3 ± 501.3

133.4 ± 179.2

400

475.8 ± 517.9

258.2 ± 213.5

Week 24

100

95.7 ± 115.6

65.0 ± 80.7

200

220.6 ± 253.3

118.1 ± 158.5

300

449.0 ± 291.7

139.4 ± 121.6*

400

736.0 ± 701.5

328.1 ± 402.6

  1. P < 0.05, vs. SWS-4000 at the same flow rate