From: In vivo and in vitro performance of a China-made hemodialysis machine: a multi-center prospective controlled study
Test time
Heparin pump (mL/h)
Actual error (ml)
SWS-4000
Fresenius 4008
Week 1
2
0.0
4
0.04 ± 0.13
0.03 ± 0.08
Week 12
0.04 ± 0.07
0.01 ± 0.05
Week 23
0.10 ± 0.21
0.13 ± 0.23
0.08 ± 0.19