Skip to main content

Table 3 Ultrafiltration errors for the two types of dialysis machines at different time points under different ultrafiltration rate settings

From: In vivo and in vitro performance of a China-made hemodialysis machine: a multi-center prospective controlled study

Test time

Ultrafiltration rate (mL/h)

Actual error (ml)

SWS-4000

Fresenius 4008

Week 1

200

7.9 ± 12.0

6.6 ± 8.6

400

10.5 ± 17.9

4.4 ± 6.4

600

4.9 ± 5.4

6.4 ± 10.0

1000

20.5 ± 23.2

14.1 ± 3.8

Week 12

200

8.2 ± 11.8

6.8 ± 9.6

400

8.4 ± 12.5

4.9 ± 9.7

600

16.6 ± 27.5

8.8 ± 12.5

1000

14.1 ± 15.4

14.6 ± 14.0

Week 23

200

10.7 ± 14.5

10.6 ± 13.7

400

16.0 ± 20.0

6.6 ± 13.2

600

11.3 ± 13.2

8.8 ± 13.7

1000

16.8 ± 16.3

25.1 ± 27.9