Skip to main content

Table 1 Comparison of changes in Kt/V and URR values after dialysis between the two types of dialysis machines

From: In vivo and in vitro performance of a China-made hemodialysis machine: a multi-center prospective controlled study

 

SWS-4000

Fresenius 4008

Week 1

Week 4

Week 1

Week 4

Kt/V

1.34 ± 0.49

1.39 ± 0.66

1.36 ± 0.44

1.42 ± 0.64

URR

0.66 ± 0.08

0.66 ± 0.95

0.66 ± 0.08

0.67 ± 0.09