Table 3 Summary of Bland–Altman analysis for each pigs

From: Improved pressure contour analysis for estimating cardiac stroke volume using pulse wave velocity measurement

Pig no.

Bland–Altman results (ml)

Precision (%)

Pig 1

−6.8 [−18.9 to 3.0]

31

Pig 2

1.7 [−1.8 to 5.7]

20

Pig 3

−3.3 [−10.5 to 3.6]

22

Pig 4

−2.8 [−12.5 to 5.4]

30

  1. Data are presented as the bias [2.5th–97.5th percentiles], where bias is the mean difference between measured and estimated SV. Precision is calculated as half of 95% range divided by mean SV for each pig