From: Analytical method for calculation of deviations from intended dosages during multi-infusion

Fig. 8

Flow rate \(u_{\mathcal {M}}^{(R)diff}(\tau )\) as function of time \(\tau \) for different values of the resistance R\(_{cath}\) and the mechanical compliance C\(_{red}\), according to the theoretical model. (1) \(C_{red}=C_{standard}, R_{cath}=3 R_{standard}.\) (2) \(C_{red}=C_{standard}, R_{cath}=2 R_{standard}.\) (3) \(C_{red}=C_{standard}, R_{cath}=R_{standard}\). (4) \(C_{red}= 0.52 C_{standard}, R_{cath}=R_{standard}\). (5) \(C_{red}= 0.36 C_{standard}, R_{cath}=R_{standard}\)