Skip to main content

Table 2 Percentage differences in the hemodynamic parameters of the parent vessel between the three different methods

From: Phantom-based experimental validation of fast virtual deployment of self-expandable stents for cerebral aneurysms

 

D1 (%)

D2 (%)

Average of velocity on aneurysm (m/s)

9.52

9.17

Area average of WSS on aneurysm (Pa)

1.26

1.95

Area average of WSS on parent vessel (Pa)

4.97

2.83

  1. D1 = |(Y-X)|/X; D2 = |(Z-X)|/X; X, Y, and Z represent corresponding hemodynamic parameters from the realistic phantom, FVS simulation, and FEM simulation respectively. WSS wall shear stress