Skip to main content

Table 1 Hemodynamic parameters of the parent vessel in the three different methods

From: Phantom-based experimental validation of fast virtual deployment of self-expandable stents for cerebral aneurysms

 

Average of velocity inside aneurismal sac (m/s)

Area average of WSS on the aneurysm (Pa)

Area average of WSS on the parent vessel (Pa)

Realistic (X)

0.0829

0.8317

4.1118

FVS (Y)

0.0908

0.8212

3.9075

FEM (Z)

0.0905

0.8155

4.2283

  1. X corresponding hemodynamic parameters in realistic phantom deployment of Enterprise stent; Y corresponding hemodynamic parameters in FVS simulation of Enterprise stent; Z corresponding hemodynamic parameters in FEM simulation of Enterprise stent; WSS wall shear stress