From: Correcting motion in multiplanar cardiac magnetic resonance images
Patient | Two chamber view | Three chamber view | Four chamber view | ||||||
---|---|---|---|---|---|---|---|---|---|
 | H (mm) | D | J | H (mm) | D | J | H (mm) | D | J |
1 before | 12.74 | 0.89 | 0.80 | 9.52 | 0.89 | 0.81 | 9.29 | 0.95 | 0.91 |
1 after | 12.35 | 0.97 | 0.94 | 11.70 | 0.92 | 0.85 | 7.59 | 0.94 | 0.89 |
2 before | 9.52 | 0.95 | 0.91 | 17.44 | 0.92 | 0.85 | 11.61 | 0.85 | 0.74 |
2 after | 11.09 | 0.97 | 0.93 | 14.20 | 0.88 | 0.79 | 12.79 | 0.96 | 0.93 |
3 before | 7.97 | 0.85 | 0.74 | \(\infty\) | 0.00 | 0.00 | 18.52 | 0.73 | 0.57 |
3 after | 6.93 | 0.84 | 0.72 | 28.37 | 0.66 | 0.49 | 16.90 | 0.89 | 0.80 |
4 before | 9.34 | 0.87 | 0.77 | 20.87 | 0.93 | 0.87 | 13.15 | 0.87 | 0.76 |
4 after | 5.16 | 0.94 | 0.88 | 3.86 | 0.93 | 0.88 | 4.12 | 0.93 | 0.87 |
5 before | 17.88 | 0.82 | 0.70 | 13.96 | 0.95 | 0.91 | 12.25 | 0.95 | 0.91 |
5 after | 5.71 | 0.98 | 0.97 | 9.94 | 0.97 | 0.94 | 7.86 | 0.98 | 0.95 |