Table 5 Detailed results for each patient for part 2 of the study

From: Interference between surgical magnetic drapes and pacemakers: an observational study comparing commercially available devices and a new magnetically isolated drape

Subject

Age

Gender

Weight

(kg)

Height

(m)

BMI

Pacemaker model

Asynchronous rhythm with drape

Prototype

JCM

LDR

SDR

CVD

1

74

M

89

1.78

28.1

Medtronic

No

No

No

No

No

2

87

F

44

1.52

19

Medtronic

No

No

No

Yes

Yes

3

66

M

89

1.73

29.7

Medtronic

No

No

No

No

Yes

4

78

M

70

1.70

24.2

Medtronic

No

No

No

No

Yes

5

86

F

52

1.60

20.3

Medtronic

No

No

No

No

Yes

6

87

F

66

1.55

27.5

Medtronic

No

No

No

No

Yes

7

88

F

61

1.65

22.4

Medtronic

No

No

No

No

Yes

8

81

M

77

1.80

23.8

Medtronic

No

Yes

No

No

Yes

9

69

M

80

1.60

31.2

Boston Scientific

No

No

No

No

No

10

75

M

60

1,49

28.6

Boston Scientific

No

Yes

No

No

Yes

11

64

F

94

1.60

36.7

Boston Scientific

No

No

No

No

No

12

92

F

46

1,49

21.9

Boston Scientific

No

Yes

No

No

Yes

13

78

F

89

1.70

30.8

Boston Scientific

No

No

No

No

Yes

14

76

M

91

1.75

29.7

Boston Scientific

No

No

No

No

No

15

83

M

95

1.78

30

Boston Scientific

No

Yes

No

No

Yes

16

79

F

79

1.68

28

Boston Scientific

No

No

No

No

Yes

17

72

M

75

1.75

24.5

Medtronic

No

No

No

No

Yes

18

80

F

63

1.60

24.6

Boston Scientific

No

No

No

No

Yes

19

72

F

66

1.52

28.6

Medtronic

No

No

No

No

Yes

20

77

M

70

1.65

25.7

Medtronic

No

Yes

No

No

Yes