Skip to main content

Table 3 Results from part 2 of the study (n = 20)

From: Interference between surgical magnetic drapes and pacemakers: an observational study comparing commercially available devices and a new magnetically isolated drape

Drapes tested

Prototype

CVD

JCM

SDR

LDR

Patients who demonstrated asynchronous rhythm (%)

N = 0

N = 16 (80 %)

N = 5 (25 %)

N = 1 (5 %)

N = 0

Age (years) (mean ± SD)

N/A

80 ± 7

82 ± 7

87

N/A

Gender

N/A

M 7/F 9

M 4/F 1

M 0/F 1

N/A

Weight (kg) (mean ± SD)

N/A

69 ± 15

70 ± 18

44

N/A

Height (m) (mean ± SD)

N/A

1.66 ± 0.09

1.74 ± 0.08

1.52

N/A

Body mass index (kg/m2) (mean ± SD)

N/A

26 ± 4

26 ± 3

19

N/A

Pacemaker (Boston Scientific/Medtronic)

N/A

6/10

3/2

0/1

N/A