Table 1 Number of accumulated AE signals, mean value and implant failure mode under fatigue test of all endodontically treated TISP samples

From: Biomechanical interactions of endodontically treated tooth implant-supported prosthesis under fatigue test with acoustic emission monitoring

 

TIS (2-piece) implant

RS (3-piece) implant

Number of accumulated AE signals

4084

565

940

226134

130834

335

286

412

628

522

Mean ± SD

1863 ± 1578

437 ± 125

Implant failure

Abutment screw fracture

No damage was noted

Abutment screw fracture

Abutment screw fracture

Abutment screw fracture

No damage was noted

No damage was noted

No damage was noted

No damage was noted

No damage was noted