Table 8 Bias of VDW measurement

From: Wound perimeter, area, and volume measurement based on laser 3D and color acquisition

Wound

Before

After

Change

Expected

Measured

Rel. diff.

Expected

Measured

Rel. diff.

Expected

Measured

Rel. diff.

1

1507

1428

−5.2%

992

936

−5.6%

515

492

−4.5%

2

2772

2715

−2.1%

1602

1589

−0.8%

1171

1126

−3.8%

3

2688

2716

1.1%

1565

1528

−2.4%

1122

1188

5.8%

  1. Measured VDW values compared to the expected values. All values are in mm3 except where noted differently.