Table 1 The oxygen content within the channel vs. the time series of the numbered spectra that presented in Figure 6

From: A lab-on-a-chip for hypoxic patch clamp measurements combined with optical tweezers and spectroscopy- first investigations of single biological cells

Absorption spectra (No. in Figure 6) Time (Seconds) O 2 -Concentrations (%)
1 0 17.83 ± 0.14
2 27 ± 0.5 13.030 ± 0.100
3 83 ± 0.5 10.720 ± 0.180
4 128 ± 1 9.607 ± 0.016
5 195 ± 1 6.430 ± 0.030
6 191 ± 1 4.465 ± 0.026
7 227 ± 1 3.481 ± 0.014
8 255 ± 1 2.941 ± 0.018
9 282 ± 1 0.103 ± 0.006
10 298 ± 1 0.017 ± 0.016
11 5400 ± 2 0.023 ± 0.008
12 5828 ± 2 13.650 ± 0.160