From: Reduction of CPR artifacts in the ventricular fibrillation ECG by coherent line removal

Figure 3

Windowed Fourier transform with the same basic ECG as used in Fig 2, but now for the VF-ECG purged from CPR-artifacts by coherent line removal and with a frequency window of [2 Hz, 30 Hz] for determination of mean, median, dominant and 95%-edge frequency. The CPR-related artifacts are barely visible in the spectrogram. The mean and median frequency follow the "fibrillation band" rather well, apart from the last stage of the experiment (> 1100 sec), where no "fibrillation band" is visible any more (probably due to asystole). The lower panel indicates that the relative power in the frequency window [0.33 Hz, 2.2 Hz] as compared to the frequency window [0.33, 30 Hz] is much smaller now than in Fig 2.