Table 2 E-ratio for Large probe vs. Small probe

From: In-vitro mapping of E-fields induced near pacemaker leads by simulated MR gradient fields

Lead

E-ratio Large probe

E-ratio Small probe

Small probe vs. Large probe

Simulated lead (insulated wire)

3.6

9.5

2.6

Active fixation tip

15.3

22.0

1.6

Tined Tip

18.7

30.4

1.6

  1. Note: See the measurement uncertainty issues" section and the "Alteration of intended therapy by magnetically-induced E-fields" section.