Skip to main content
Figure 7 | BioMedical Engineering OnLine

Figure 7

From: An e-learning application on electrochemotherapy

Figure 7

Local electric field distribution within a model of cutaneous (protruding) tumour. Local electric field distribution inside the protruding tumour model for two different applied voltages on the electrodes: a) U = 300 V and b) U = 600 V. The electrodes are 4 mm wide and 4 mm apart in both cases. Electric field distribution inside the models for two distances between electrodes: c) d = 4 mm and d) d = 8 mm. The electrodes are 4 mm wide with the applied voltage U = 600 V in both cases.

Back to article page