Table 4 Effect of stent-graft Youngs Modulus

From: Fluid-structure interaction of a patient-specific abdominal aortic aneurysm treated with an endovascular stent-graft

Stent-graft Young's Modulus (MPa)

Peak drag force (N)

Peak aneurysm wall stress (MPa)

Compliance at proximal neck (1 × 10-4/mmHg)

Compliance at maximum diameter (1 × 10-4/mmHg)

5

4.86

0.061

6.79

0.92

10

4.85

0.038

5.85

0.46

15

4.83

0.019

4.81

0.33

  1. Table detailing the sensitivity of the results to a change in stent-graft stiffness.