Skip to main content

Table 1 Features used in signal abnormality index jSQI. k is the beat index.

From: Artificial arterial blood pressure artifact models and an evaluation of a robust blood pressure and heart rate estimator

Feature

Description

Abnormality CRITERIA

P s

Systolic blood pressure (SBP)

P s > 300 mmHg

P d

Diastolic blood pressure (DBP)

P d < 20 mmHg

P m

Mean arterial pressure (MAP)

P m < 30 or P m > 200 mmHg

f

Instantaneous heart rate (60/T)

f < 20 or f > 200 BPM

P p

Pulse pressure (P s - P d )

P p < 20 mmHg

w

Noise term: mean of negative slopes

w < -40/100 mmHg/ms

P s [k] - P s [k - 1]

Absolute change in instantaneous SBP

|ΔP s | > 20 mmHg

P d [k] - P d [k - 1]

Absolute change in instantaneous DBP

|ΔP d | > 20 mmHg

T[k] - T[k - 1]

Absolute change in instantaneous HR

|ΔT| > 2/3 s