Skip to main content

Table 1 Measured Potential Field Gradient at Defibrillation Threshold

From: Extended charge banking model of dual path shocks for implantable cardioverter defibrillators

 

Lowest Gradient (V/cm)

Highest Gradient (V/cm)

Average Gradient (V/cm)

Tang et al.[66]

2

37

11.7

Witkowski et al.[67]

4.2

30.6

12.9

Tang et al.[29]

3.5–5.8

49.1–104.0

Not given

Chen et al.[68]

7 to 8

99 to 153

Not given