Overview of the Extended Charge Banking Model for dual-path defibrillation. (A) The Charge Banking Model estimates the response of the heart to a shock as an RC circuit model. (B) The effect of a shock on many small sections of the heart may be estimated by modeling the response of each section as a separate RC circuit. (C) A weighting function approximating the current distribution through the heart for single path and dual path shocks is combined with the sectioned heart model to predict the response of the whole heart to a shock. See Figures 3 and 4 for additional labels and text for explanation. (D) Successful defibrillation is predicted by determining the effectiveness of single and dual path shocks in causing a minimum threshold model voltage in a critical mass of the heart. See Figure 5 for coordinate labels and text for further explanation.