Table 5 Normalization regression parameters. This gives the logarithmic response curve fitted parameter Molybdenum/Molybdenum (Mo/Mo) distribution quantities for both tissue types using the beam-current readout normalization (mAs). Acquisition mAs system values are indicated by the subscript of the left column entries.

From: Effective x-ray attenuation coefficient measurements from two full field digital mammography systems for data calibration applications

Mo/Mo 26 kVp

μ

σ

Δ

l

σl

R2

Fat160

0.568

0.002

0.005

5.054

0.01

0.999

 

0.550–0.586

  

4.97–5.13

  

Fat110

0.569

0.002

0.006

5.058

0.010

0.999

 

0.549–0.590

  

4.97–5.14

  

Gland160

0.827

0.002

0.016

4.851

0.010

0.998

 

0.777–0.878

  

4.63–5.07

  

Gland200

0.826

0.002

0.015

4.853

0.010

0.998

 

0.777–0.875

  

4.64–5.06

 Â