Table 4 Effective attenuation coefficient comparisons. This shows the numerically derived attenuation coefficients, μT, the associated regression variation, Δ, and R2 values for the Mo/Mo, Mo/Rh, and Rh/Rh configurations for the same voltage settings (kVp) used for the phantom imaging. Attenuation coefficient confidence intervals are cited beneath the respective values. The <μ > column gives the averaged FL + KS measured attenuation coefficients for the respective tissue types. Percent differences between < μ > and μ T are listed parenthetically next to the respective μ T values. Other formatting is the same used above.

From: Effective x-ray attenuation coefficient measurements from two full field digital mammography systems for data calibration applications

Mo/Mo

μ T

Δ

R2

<μ >

Fat26

0.595 (2.8)

0.010

0.999

0.578

 

0.575–0.613

   

Gland26

0.871 (4.1)

0.020

0.998

0.835

 

0.818–0.923

   

Fat27

0.585 (2.9)

0.010

0.999

0.565

 

0.562–0.604

   

Gland27

0.836 (3.4)

0.021

0.997

0.807

 

0.777–0.894

   

Mo/Rh

    

Fat28

0.529 (4.0)

0.010

0.999

0.508

 

0.511–0.547

   

Gland28

0.764 (6.8)

0.014

0.998

0.713

 

0.772–0.807

   

Fat29

0.516 (3.8)

0.010

0.999

0.498

 

0.498–0.535

   

Gland29

0.735 (5.3)

0.020

0.998

0.696

 

0.690–0.779

   

Rh/Rh

    

Fat30

0.462 (4.3)

0.010

0.999

0.442

 

0.448–0.476

   

Gland30

0.675 (5.7)

0.010

0.999

0.636

 

0.647–0.703