Table 6 Quantification of the electric field strength for in-homogeneousmodels – calculated Ett min , Ett max and E max parameters.

From: Analytical and numerical quantification and comparison of the local electric field in the tissue for different electrode configurations

Needle electrode configuration U/d = 1.15 V/cm U/d = 1300 V/cm
  Target tissue
Ettmax(V/cm)
Target tissue
Ettmin(V/cm)
Entire tissue
Emax(V/cm)
Target tissue
Ettmax(V/cm)
Target tissue
Ettmin(V/cm)
Entire tissue
Emax(V/cm)
2 (Fig. 6a) 0.558 0.364 7.001 628.05 409.9 7887.1
4 (Fig. 6b) 0.572 0.539 6.104 643.4 606.2 6872.5
6 (Fig. 6c) 0.741 0.639 5.339 833.9 718.9 6011.2
  1. Quantification of the electric field strength (E) for in-homogeneous models with needle electrode configurations which are defined in Fig. 2a-c. Conductivity of the target tissue is σtt = 0.4 S/m and the conductivity of the surrounding tissue σst = 0.2 S/m. Calculated minimal E (Ett min ) and maximal E (Ett max ) inside the target tissue and maximal E within the entire tissue E max for different needle electrode configurations. The results for models shown in Figs. 6a-6c were calculated for applied voltage U = 1 V ( d = 5 3  mm MathType@MTEF@5@5@+=feaafiart1ev1aaatCvAUfKttLearuWrP9MDH5MBPbIqV92AaeXatLxBI9gBaebbnrfifHhDYfgasaacH8akY=wiFfYdH8Gipec8Eeeu0xXdbba9frFj0=OqFfea0dXdd9vqai=hGuQ8kuc9pgc9s8qqaq=dirpe0xb9q8qiLsFr0=vr0=vr0dc8meaabaqaciaacaGaaeqabaqabeGadaaakeaacqWGKbazcqGH9aqpcqaI1aqndaGcaaqaaiabiodaZaWcbeaakiabbccaGiabb2gaTjabb2gaTbaa@349D@ ). Furthermore, by scaling the results we also calculated the parameters Ett min , Ett max and E max for U/d = 1300 V/cm.