Table 2 Quantification of electric field strength for plate electrodes models – calculated Ett min , Ett max and E max parameters.

From: Analytical and numerical quantification and comparison of the local electric field in the tissue for different electrode configurations

2 plate electrode configuration U/d = 1.15 V/cm U/d = 1300 V/cm
  Target tissue
Ettmax(V/cm)
Target tissue
Ettmin(V/cm)
Entire tissue
Emax(V/cm)
Target tissue
Ettmax(V/cm)
Target tissue
Ettmin(V/cm)
Entire tissue
Emax(V/cm)
(Fig. 3a) U/d = 1.15 U/d = 1.15 U/d = 1.15 U/d = 1300 U/d = 1300 U/d = 1300
(Fig. 3b) 1.152 1.113 5.515 1297.0 1253.0 6209.0
(Fig. 3c) 1.081 0.691 5.533 1217.0 777.9 6229.2
  1. Calculated minimal E (Ett min ) and maximal E (Ett max ) inside the target tissue and maximal E within the entire tissue E max for different plate electrode configurations as defined in Fig. 1 are given. The results are given for applied voltage U = 1 V giving U/d = 1.15 V/cm and for applied voltage (scaled results) for U = 1125.83 V giving U/d = 1300 V/cm.