Table 5 1:1 Pulsatile Flow Results for Repeat Measurements in Standing Position.

From: Physiologically-based testing system for the mechanical characterization of prosthetic vein valves

Test Condition Test Run V R MathType@MTEF@5@5@+=feaafiart1ev1aaatCvAUfKttLearuWrP9MDH5MBPbIqV92AaeXatLxBI9gBaebbnrfifHhDYfgasaacH8akY=wiFfYdH8Gipec8Eeeu0xXdbba9frFj0=OqFfea0dXdd9vqai=hGuQ8kuc9pgc9s8qqaq=dirpe0xb9q8qiLsFr0=vr0=vr0dc8meaabaqaciaacaGaaeqabaqabeGadaaakeaadaajbaqaaGqabiab=zfawbaadaWgaaWcbaGae8Nuaifabeaaaaa@2F65@ (cm3) %Ref (%) REL (mJ) η(%)
STBR 1st 5.83 ± 0.688 18.1± 2.12 2.73 ± 0.654 36.7 ± 0.785
  2nd 3.99 ± 0.537* 12.7 ± 2.07* 1.29 ± 0.371* 38.4 ± 0.52*
  Change 1.84 5.4 1.44 1.7
STAF 1st 2.21 ± 0.978 7.10 ± 3.27 0.451 ± 0.411 99.4 ± 0.562
  2nd 1.55 ± 1.74 4.75 ± 5.27 0.384 ± 0.375 99.6 ± 0.443
  Change 0.66 2.35 0.067 0.2
  1. * p < 0.05 vs. 1st Test Run