Table 3 1:1 Pulsatile Flow Results for Prototype vs. 'No Device' in Standing Position.

From: Physiologically-based testing system for the mechanical characterization of prosthetic vein valves

Test Condition Device V R MathType@MTEF@5@5@+=feaafiart1ev1aaatCvAUfKttLearuWrP9MDH5MBPbIqV92AaeXatLxBI9gBaebbnrfifHhDYfgasaacH8akY=wiFfYdH8Gipec8Eeeu0xXdbba9frFj0=OqFfea0dXdd9vqai=hGuQ8kuc9pgc9s8qqaq=dirpe0xb9q8qiLsFr0=vr0=vr0dc8meaabaqaciaacaGaaeqabaqabeGadaaakeaadaajbaqaaGqabiab=zfawbaadaWgaaWcbaGae8Nuaifabeaaaaa@2F65@ (cm3) %Ref (%) REL (mJ) η(%)
STBR None 59.6 ± 0.830 205 ± 5.37 283 ± 7.93 -319 ± 22.1
  Prototype 5.83 ± 0.688 18.1 ± 2.12 2.73 ± 0.654 36.7 ± 0.785
STAF None 7.35 ± 0.28 24.6 ± 0.859 4.30 ± 0.333 93.9 ± 0.429
  Prototype 2.21 ± 0.978 7.10 ± 3.27 0.451 ± 0.411 99.4 ± 0.562