Skip to main content

Table 4 The averages and standard deviations of h at different flow rates

From: In vitro calibration of a system for measurement of in vivo convective heat transfer coefficient in animals

Flow (L/min) Distilled water (W/(m2·K)) Tap water (W/(m2·K)) Saline (W/(m2·K))
  Average SD Average SD Average SD
0.0 5100 500 5500 400 5400 400
1.0 9000 1200 8000 1100 8400 800
2.0 10100 1800 9500 1100 9700 1100
3.0 12000 1800 10300 800 10300 1400
4.0 12200 1900 11000 400 12400 1800
5.0 13000 1800 12300 800 13600 1700