Figure 14 | BioMedical Engineering OnLine

Figure 14

From: Finite volume analysis of temperature effects induced by active MRI implants: 2. Defects on active MRI implants causing hot spots

Figure 14

Temporal development of the critical volume for 200 mW hot spot power loss for various distances between the hot spot and a planar heat sink simulating blood flow. The maximum critical volume depends on the distance between the hot spot and the heat sink simulating the blood flow inside a large vessel. For a very short distance of only a few cell layers, the critical volume nevertheless reaches more than 1 mm3. Increasing the distance strongly increases the critical volume. The four graphs show the fast buildup of the critical volume within a few seconds.