Figure 11

From: Finite volume analysis of temperature effects induced by active MRI implants: 2. Defects on active MRI implants causing hot spots

Figure 11

Various wire radii. An increasing wire radius with a larger capability of heat transport along the wire improves the heat distribution in axial direction with the same consequences as in figure 8. With increasing wire radius the critical volume becomes smaller, but it stays unacceptably large even for an unrealistic wire radius of 0.5 mm (1 mm diameter).