Table 3 Experimental: number of data points within +/- 10, 50 and 100 % of the target value for each of the three chambers tested. The regions of interest used are depicted in Figures 10-12.

From: The imperative for controlled mechanical stresses in unraveling cellular mechanisms of mechanotransduction

Chamber Within ±10% Within ±25% Within ±50%
Oligene FCS 0% 0% 26.92%
# data points/total (0/26) (0/26) (7/26)
Bioptechs FCS2 20.00% 73.33% 96.67%
# data points/total (6/30) (22/30) (29/30)
Warner RC-30HVS 21.43% 39.29% 85.71%